A patient is moving in a magnetic field for 2 sec. and experiences a magnetic spatial gradient of 200G/cm. What dB/dt are they experiencing?
100T/s
Below the IEC limit
400T/s
10T/s
Below the IEC limit
- / 4
Page 2 of 227
The limit of dB/dt according to the IEC 60601-2-33 document for movement in a static magnetic field is 3T/s. 200G/cm is = 2T/m and 2T/m divided by 2 seconds = 1T/m/s
This would be below the 3T/s IEC limit
You’re scanning a brain with a resonant frequency of 128MHz.Your MR conditions state to only scan with a 1.5T unit in normal mode. Can you meet the MR conditions?
Yes
No
No. (128MHz represent a 3T unit.)
Which of the following are true of a department according to the ACR Manual on MRI Safety 2020? Select all that apply. 2 / 4
Page 3 of 227
Patients are screened for threats in Zone 3
Ferrodetectors should be at the entrance of Zone 3
The entrance to zone 3 should have a keypad/badge system to control access
MR screening forms should be filled out in zone 4
Ferrodetectors should be at the entrance of Zone 3
The entrance to zone 3 should have a keypad/badge system to control access
Due to induced electrical fields produced in our patients due to flowing blood and our static magnetic field, what can be demonstrated?
Magnetophosphenes 3 / 4
Page 4 of 227
All of the following
Nystagmus
Magnetohydrodynamic effect
Magnetohydrodynamic effect
Given the following parameters, which has the lowest heating potential when dealing with TR, Pulse Duration, and flip angle?
TR= 2000 Pulse Duration=1ms Flip angle = 100
TR= 1500 Pulse Duration=2ms Flip angle = 160
TR= 2000 Pulse Duration=2ms Flip angle = 180
TR= 1500 Pulse Duration=1ms Flip angle = 180