ACCU-CHEK INFORM II GLUCOMETER
COMPETENCY TEST, VA WITH
COMPLETE SOLUTION
Question : Quality control testing must be performed:
- Each 24 hours of meter use
- Each time a new bottle of test strips is opened
- When the meter has been dropped or the battery
- When test results contradict clinical symptoms
- All of the above
replaced
Correct answer: All of the above
Question : Critical values require immediate attention.
What is the range for critical values on the glucometer?
- <65 or >100 mg/dL
- <50 or >500 mg/dL
- 0 or >1000 mg/dL
Correct answer: <50 or >500 mg/dL
- | Page
Question : Operator IDs allow only properly trained
operators to use the glucometer. Is it ever acceptable for operators to share their ID or allow others to do testing under their account?
- Yes
- Absolutely not
Correct answer: Absolutely not
Question : When does the meter need to be cleaned?
- Once every 24 hours of meter use
- Each shift
- Before and after each patient or after running the quality
control.
Correct answer: Before and after each patient or after
running the quality control.
Question : Quality control bottles need to be dated with
the expiration date. How long are the quality controls stable after bottles are opened?
- | Page
- 3 months or the date on the bottle. Whichever comes
- 6 months or the date on the bottle. Whichever comes
- 1 year
first.
first.
Correct answer: 3 months or the date on the bottle.
Whichever comes first.
Question : What should you use to clean the glucometer?
- Clorox or Super Sani Germicidal Wipes
- CaviWipes
Correct answer: Clorox or Super Sani Germicidal Wipes
Question : During situations of decreased peripheral
blood flow (such as severe dehydration, shock or
peripheral vascular disease.):
- It is still acceptable to use the glucometer to obtain a
- A venous blood sample should be drawn and brought to
glucose result
the Laboratory for testing.
- | Page
Correct answer: A venous blood sample should be drawn
and brought to the Laboratory for testing.
Question : Prior to any patient testing, the operator needs
to verify the patient's full name and full social security number.
- True
- False
Correct answer: True
Question : A comment of "Provider notified" or "Rx as
needed" is required with EVERY critical result.
- True
- False
Correct answer: True
Question : Can the meter be used on critically ill patients?
Critically ill patients are patients that are on mechanical ventilation with a mean arterial pressure <60 mmHg despite the use of pressors.
60>50>50>65>