2024 ABRET Practice Test New Full Test | Questions and Answers ( Included ) 100% Correct

  • in which type of evoked potentials is the scalp position of recording electrodes
  • most critical?

Answer: lower extremity SSEPs

  • if an evoked potential has a voltage of 10 uV and the associated noise has a
  • voltage of 50 uV, what is the signal to noise ratio after 100 averaged responses?Answer: 2:1

  • to display median nerve SSEP responses as an upward deflection, which of the
  • following must be known?

Answer: waveform polarity and polarity convention of the amplifier inputs

  • if an amplifier is rated as having 120 dB common mode rejection ratio, which of
  • the following ratios in phase to out of phase signals would be expected?Answer: 1,000,000:1

  • which of the following cranial nerves innervates an extraocular muscles

Answer: - abducens

  • Which of the following is an example of an SSEP far-field subcortical poten tial?

Answer: N34

  • what is the segmental motor innervation of the peroneus longus?

Answer: L4, L5, S1

  • which of the following opposes any change in current

Answer: inductance

  • the dorsal column pathyways synapse at which loci on their way to the sensory
  • cortex

Answer: nucleus cuneatus and thalamus

  • with referential recording, maximal amplitude cortical evoked potentials
  • following stimulation of the left posterior tibial nerve usually would be found between

Answer: C3 and CZ

  • / 3
  • which area of the spinal cord is most likely to have the poorest blood supply

Answer: thoracic

12. the posterior longitudinal ligament lies within the vertebral canal:

Answer: in front of the vertebral body

  • which of the following would result in the best resolution in the ana-
  • log-to-digital conversion process?

Answer: the maximal peak-to-peak signal amplitude should match the converter

range

  • an increase in stimulation intensity causes a linear increase in which of the
  • following SSEP components

Answer: P31

  • epidural electrodes are made of platinum because

Answer: other metals may be- come toxic when currents are applied

  • during a median nerve preoperative study, a patient has discomfort at the wrist.
  • stimulation electrodes are 10mm gold disk, stimulus intensity is 20 mA, and stimulus duration is 100 msec. which of the following would most likely reduce the patient's discomfort while maintaining adequate stimulation?

Answer: de- crease number of repetitions or use a larger stimulating electrode

  • what improves the signal-to-noise ratio by the square root of the number of
  • responses added?

Answer: signal averaging

  • in BAEPs, wave I can most easily be differentiated from cochlear micro-
  • phonics based upon

Answer: click polarity

  • what is the interstimulus of a 50 Hz click stimulus

Answer: 20 msec

  • SSEPs recorded from the scalp after electrical stimulation of the peripheral nerve
  • are conducted through the

Answer: dorsal column on the side of stimulation

  • monitoring the posterior tibial nerve SSEPs is most appropriate in cere-
  • brovascular procedures involving which of the following arteries?

Answer: anterior cerebral 2 / 3

  • identification of wave V of the BAEP may require

Answer: decreasing stimulus inten sity

  • in spinal surgery, the primary purpose of the popliteal fossa recording electrode
  • is to

Answer: verify adequacy of stimulus

  • while performing a posterior tibial SSEP, hypothermia of the lower extremities is
  • most likely to cause

Answer: slowed peripheral conduction velocity

  • which peak in tibial nerve SSEPS is most analogous to P14 of median nerve SSEP

Answer: P31

  • during posterior fossa surgery, when BAEPs are used for monitoring brain- stem
  • function, it is most important to

Answer: stimulate and record from the contralateral side

  • which cranial nerve is most appropriate to monitor during resection of a 4cm
  • vestibular schwannoma

Answer: facial

  • during carotid endarterectomy, what ssep modality is recommended for
  • monitoring cerebral function

Answer: median nerve

  • if the EEG is preserved during carotid endarterectomy surgery, it indicates-

Answer: adequate cortical perfusion and oxengenation

  • alternating current is used when checking electrode impedance to

Answer: avoid electrode polarization

  • which of the following is most likely to cause 60 Hz contamination of evoked
  • potential waveforms in the operating room?

Answer: fluorescent lights

  • a postincision baseline recording is used to

Answer: establish a standard for the patient

  • which of the following would most likely be normal int he pre-operative
  • baseline upper extremity SSEP of a patient with a lesion in the dorsal column?

Answer: ERB's point potential

  • / 3

Download Document

Buy This Document

$30.00 One-time purchase
Buy Now
  • Full access to this document
  • Download anytime
  • No expiration

Document Information

Category: Study Guides
Added: Aug 17, 2025
Description:

2024 ABRET Practice Test New Full Test | Questions and Answers ( Included ) 100% Correct 1. in which type of evoked potentials is the scalp position of recording electrodes most critical? Answer: l...

Get this document $30.00